Jove
Visualize
联系我们
JoVE
x logofacebook logolinkedin logoyoutube logo
关于 JoVE
概览领导团队博客JoVE 帮助中心
作者
出版流程编辑委员会范围与政策同行评审常见问题投稿
图书馆员
用户评价订阅访问资源图书馆顾问委员会常见问题
研究
JoVE JournalMethods CollectionsJoVE Encyclopedia of Experiments存档
教育
JoVE CoreJoVE BusinessJoVE Science EducationJoVE Lab Manual教师资源中心教师网站
使用条款与条件
隐私政策
政策

相关概念视频

Glassware Calibration01:11

Glassware Calibration

163
Accurate calibration of glassware, such as volumetric flasks, pipettes, and burettes, is essential to ensure accurate measurements in the analytical laboratory. Calibration helps maintain consistency across measurements and prevents errors arising from inaccurate volumes.
Volumetric flasks: Volumetric flasks are designed to prepare aqueous solutions of precise volumes accurately with a calibration line on the neck. To calibrate a volumetric flask, it is important to fill it with distilled...
163

您也可能阅读

相关文章

通过共同作者、期刊和引用图与本文相关的文章。

排序
Same author

Qualitative Research on Respiratory Care Education Topics in Published Literature: A Scoping Review.

Respiratory care·2026
Same author

Revisiting Diffusing Capacity of the Lung for Carbon Monoxide Variability: Measurement Science, Quality Assurance, and Clinical Interpretation in the Real World.

Chest·2025
Same author

Prioritising the Burden of Geographic Atrophy and Treatment Expectations: A Modified Nominal Group Technique Study with International Patient-Led Organisations.

Clinical ophthalmology (Auckland, N.Z.)·2025
Same author

Reconsidering Race and Ethnicity for Predicting Normal Lung Function in Spirometry.

Respiratory care·2025
Same author

Intestinal Mycobacterium avium Complex in a Patient With AIDS.

The American journal of gastroenterology·2024
Same author

Prediction of Hospital Readmission Using the CORE and CORE+ Scores in Persons With COPD.

Respiratory care·2024

相关实验视频

Updated: May 23, 2025

Author Spotlight: Studying Bacterial Growth in 3D Hydrogel Matrices
05:46

Author Spotlight: Studying Bacterial Growth in 3D Hydrogel Matrices

Published on: January 19, 2024

2.0K

需要进行体积线性和注射器的训练强度DLCO 检查质量控制.

Ellen A Becker1, Ellen M Moran2, Jason B Blonshine3

  • 1Dr Becker is affiliated with Rush University Medical Center, Chicago, Illinois, USA.

Respiratory care
|March 7, 2025
PubMed
概括

肺功能设备质量控制 (QC) 的故障率很高,特别是体积线性. 技术人员需要持续的培训和明确的指导方针,而不仅仅是初次会议,以确保准确的患者测试.

关键词:
DLCO 检查 检查 DLCO 检查教育教育教育教育教育教育.肺功能测试 肺功能测试质量控制 质量控制 质量控制卷积线性 卷积线性

更多相关视频

The Quantification of Injectability by Mechanical Testing
04:46

The Quantification of Injectability by Mechanical Testing

Published on: May 13, 2020

7.8K
Subconjunctival Administration of Adeno-associated Virus Vectors in Small Animal Models
06:16

Subconjunctival Administration of Adeno-associated Virus Vectors in Small Animal Models

Published on: March 16, 2022

2.8K

相关实验视频

Last Updated: May 23, 2025

Author Spotlight: Studying Bacterial Growth in 3D Hydrogel Matrices
05:46

Author Spotlight: Studying Bacterial Growth in 3D Hydrogel Matrices

Published on: January 19, 2024

2.0K
The Quantification of Injectability by Mechanical Testing
04:46

The Quantification of Injectability by Mechanical Testing

Published on: May 13, 2020

7.8K
Subconjunctival Administration of Adeno-associated Virus Vectors in Small Animal Models
06:16

Subconjunctival Administration of Adeno-associated Virus Vectors in Small Animal Models

Published on: March 16, 2022

2.8K

科学领域:

  • 肺部医学 肺部医学
  • 医学诊断 医学诊断 医学诊断
  • 质量控制 质量控制 质量控制

背景情况:

  • 精确的肺功能测试依赖于精确的设备校准.
  • 技术人员必须在对患者进行测试之前确保肺功能设备的准确性和精度.
  • 螺旋计和扩散能力测试中的质量控制 (QC) 失败会影响患者的护理.

研究的目的:

  • 确定螺旋计体积线性和注射器肺部一氧化碳 (DLCO) QC的扩散能力的故障来源和不可接受的提交率.
  • 探索培训和第一次QC提交之间的时间对整体不可接受的提交率的影响.

主要方法:

  • 从2014年10月到2018年5月,对114个实验室进行了全球横截面研究.
  • 分析了5083个体积线性和4992个注射器DLCO检查QC提交.
  • 描述性统计和斯皮尔曼的rho相关性来评估失败率和培训时间表.

主要成果:

  • 不可接受的提交率为体积线性10%和注射器DLCO的6.2%.
  • 卷线性故障在程序和设备问题 (5.8%对6.2%) 中均分.
  • 注射器DLCO故障主要与设备相关 (5.9%对0.78%).
  • 培训后早期的质量控制失败与两个测试的总体不可接受率更高相关 (r = -0.41和r = -0.38).

结论:

  • 对于技术人员来说,一个QC培训课程是不够的.
  • 清晰的指导方针和专家跟进对于解释质量控制结果和解决设备问题至关重要.
  • 持续的支持是必要的,以保持在控制测试系统,并确保精确的肺功能诊断.