Related Experiment Video
Updated: Aug 17, 2026

Visualization of the Charcoal Agar Resazurin Assay for Semi-quantitative, Medium-throughput Enumeration of Mycobacteria
Published on: December 14, 2016
Mycobactericidal activity of selected disinfectants using a quantitative suspension test
P A Griffiths1, J R Babb, A P Fraise
1Hospital Infection Research Laboratory, City Hospital NHS Trust, Birmingham.
Abstract:
In this study, a quantitative suspension test carried out under both clean and dirty conditions was used to assess the activity of various instrument and environmental disinfectants against the type strain NCTC 946 and an endoscope washer disinfector isolate of Mycobacterium chelonae, Mycobacterium fortuitum NCTC 10,394, Mycobacterium tuberculosis H37 Rv NCTC 7416 and a clinical isolate of Mycobacterium avium-intracellulare (MAI). The disinfectants tested were; a chlorine releasing agent, sodium dichloroisocyanurate (NaDCC) at 1000 ppm and 10,000 ppm av Cl; chlorine dioxide at 1100 ppm av ClO2 (Tristel, MediChem International Limited); 70% industrial methylated spirits (IMS); 2% alkaline glutaraldehyde (Asep, Galan); 10% succinedialdehyde and formaldehyde mixture (Gigasept, Schulke & Mayr); 0.35% peracetic acid (NuCidex, Johnson & Johnson); and a peroxygen compound at 1% and 3% (Virkon, Antec International). Results showed that the clinical isolate of MAI was much more resistant than M. tuberculosis to all the disinfectants, while the type strains of M. chelonae and M. fortuitum were far more sensitive. The washer disinfector isolate of M. chelonae was extremely resistant to 2% alkaline activated glutaraldehyde and appeared to be slightly more resistant than the type strain to Nu-Cidex, Gigasept, Virkon and the lower concentration of NaDCC. This study has shown peracetic acid (Nu-Cidex), chlorine dioxide (Tristel), alcohol (IMS) and high concentrations of a chlorine releasing agent (NaDCC) are rapidly mycobactericidal. Glutaraldehyde, although effective, is a slow mycobactericide. Gigasept and Virkon are poor mycobactericidal agents and are not therefore recommended for instruments or spillage if mycobacteria are likely to be present.