Related Experiment Videos
Sternal acceleration ballistocardiography and arterial pressure wave analysis to determine stroke volume
W P McKay1, P H Gregson, B W McKay
1Department of Anaesthesia, University of Saskatchewan, Royal University Hospital, Saskatoon. wmckay@the.link.ca
Objective:
To describe a cardiac output measurement using a new method to derive and analyze the long-axis ballistocardiogram that is less invasive than pulmonary artery thermodilution.
Design:
Prospective physiologic study.
Setting:
Intensive care unit of The Halifax Infirmary, a teaching hospital of Dalhousie University, Halifax, NS.
Patients:
Thirty-nine patients in sinus rhythm with pulmonary artery thermodilution catheters or radial artery catheters in place. The first 30 subjects were the "learning set" and the next 9 were the "test set."
Interventions:
A small (54-g) accelerometer was taped on the patient's chest.
Outcome Measures:
Measurements of time and amplitude coordinates of the acceleration and radial artery pressure wavepeaks, as well as anthropometric information.
Results:
A stroke volume prediction equation was generated (R2 = 0.76) from the learning set. This equation was applied to the test set and correlated with the pulmonary artery thermodilution-derived stroke volumes (R = 0.79). Stroke volumes were compared using a previously described statistical method: a) bias (predicted > thermodilution) = 0.03 mL (95% confidence interval [CI] -4.2 to 4.8 mL); b) lower limit of agreement = -21 mL (95% CI -29 to -13 mL); c) upper limit of agreement = 22 mL (95% CI 14 to 29 mL). Of derived stroke volumes, 82% were within 15 mL of pulmonary artery thermodilution-derived values.
Conclusions:
The sternal acceleration ballistocardiogram combined with hemodynamic and demographic data in a probabilistic model shows promise of providing a less invasive measure of cardiac output than thermodilution.