UK audit of relative renal function measurement using DMSA scintigraphy
J S Fleming1, P S Cosgriff, A S Houston
1Nuclear Medicine Software Working Party, Institute of Physics and Engineering in Medicine, York, UK.
Nuclear Medicine Communications
|May 11, 1999
Summary
This study assessed variations in relative renal function measurements from DMSA scintigraphy across UK hospitals. Results show the procedure is reliable, but standardizing analysis techniques could improve consistency.
Area of Science:
- Nuclear Medicine
- Medical Imaging Analysis
- Clinical Audit
Background:
- Quantitative parameters from radionuclide imaging require consistent software analysis.
- Understanding inter-hospital variability is crucial for clinical audit of imaging techniques.
Purpose of the Study:
- To investigate variations in relative renal function measurements from static DMSA scintigraphy within the UK.
- To identify factors contributing to discrepancies in DMSA scintigraphy results.
Main Methods:
- Ten DMSA scintigraphy studies were distributed digitally to 100 UK hospitals.
- Participating hospitals analyzed studies using their standard techniques and computer systems.
- Data on analysis views, background subtraction, and region definition were collected.
Main Results:
- 88% of results were within 2 percentage points of the median relative function, and 98% within 5 points.
- Statistically significant differences were linked to analysis view and background subtraction methods.
- Variations observed in geometric mean vs. posterior views and single vs. separate region background subtraction.
Conclusions:
- Relative renal function assessment using static DMSA scintigraphy is generally reliable in the UK.
- Standardizing DMSA scintigraphy analysis techniques, particularly view selection and background subtraction, could enhance procedural consistency.
- Further standardization can improve the clinical audit of radionuclide imaging.