In vivo evaluation of the adjustable temporary venous spring filter and the RF02 temporary filter: comparative study
1Section for Experimental Radiology, National Hospital, University of Oslo, Norway.
Academic Radiology
|June 22, 1999
Summary
The spring filter demonstrated efficient clot trapping with a lower rate of caval occlusion compared to the RF02 filter in an animal model. Further development is needed for optimal performance.
Area of Science:
- Vascular surgery
- Medical device evaluation
- Inferior Vena Cava (IVC) filters
Background:
- Inferior Vena Cava (IVC) filters are crucial for preventing pulmonary embolism.
- Evaluating the in vivo performance of novel IVC filters is essential for patient safety.
- Comparing existing and new filter designs aids in clinical decision-making.
Purpose of the Study:
- To compare the in vivo efficacy of the temporary venous spring filter and the RF02 filter.
- To assess clot-trapping ability and caval occlusion rates.
- To evaluate hemodynamic changes associated with filter placement.
Main Methods:
- Ten pigs each received either a spring filter or an RF02 filter in the inferior vena cava.
- Two clots were introduced sequentially to evaluate clot-trapping capacity.
- Clot capture, caval occlusion, blood gases, and venous pressures were measured.
Main Results:
- The spring filter captured 90% of the first and 100% of the second clots.
- The RF02 filter also demonstrated consistent clot capture.
- Caval occlusion occurred less frequently with the spring filter (17%) compared to the RF02 filter (67%).
Conclusions:
- The spring filter is an effective device for clot trapping.
- The spring filter exhibited a lower rate of caval occlusion than the RF02 filter.
- Improvements in filter size and stability are recommended for the spring filter.