Extension of CadPlan algorithm to model the dose distribution under a motorized wedge
1BC Cancer Agency, Cancer Centre for the Southern Interior, Kelowna, Canada.
Physics in Medicine and Biology
|August 12, 1999
Summary
This study enhances the CadPlan treatment planning system for Elekta motorized wedges, improving off-axis ratio (OAR) accuracy. Modified algorithms now model OAR within 2.5% of measured values for improved radiation therapy.
Area of Science:
- Medical Physics
- Radiation Oncology
- Radiotherapy Treatment Planning
Background:
- The CadPlan treatment planning system uses an equivalent water thickness model for wedged fields.
- This model accurately predicts dose distribution for uniform wedges but struggles with tapered motorized wedges.
- Motorized wedges in Elekta accelerators have tapered thickness, leading to inaccuracies in the CadPlan model.
Purpose of the Study:
- To evaluate and improve the accuracy of the CadPlan treatment planning system for modeling off-axis ratios (OAR) in wedged fields from Elekta motorized accelerators.
- To address the limitations of the existing CadPlan model which assumes uniform wedge thickness.
- To achieve accurate dose distribution modeling for radiation therapy planning.
Main Methods:
- The study adapted the CadPlan algorithm by modifying effective water thickness values at off-axis distances.
- The modified algorithm was tested for 4, 6, and 18 MV beams on Elekta accelerators.
- Off-axis ratio (OAR) predictions from the modified model were compared against measured values.
Main Results:
- The original CadPlan model showed significant over/underestimations of OAR for Elekta motorized wedges (up to 10%).
- The modified CadPlan algorithm successfully modeled the OAR in the non-wedged direction.
- The enhanced model achieved accuracy within 2.5% of measured OAR values for tested beams and accelerators.
Conclusions:
- The CadPlan system requires modification to accurately model dose distributions for Elekta motorized wedges.
- Adjusting effective water thickness at off-axis positions significantly improves OAR prediction accuracy.
- The enhanced CadPlan model provides a more reliable tool for radiotherapy treatment planning with Elekta motorized wedges.