1Institute of Medical Science, Department of Anesthesiology, Faculty of Medicine, University of Tokyo, Japan.
Background:
Failed or difficult tracheal intubation is an important cause of morbidity and mortality during anesthesia. Although a number of fiberoptic devices are available to circumvent this problem, many do not allow manual control of the flexion of the tip and necessitate time-consuming preparation, special training, or the use of an external light source. To improve these limitations, the authors designed a new fiberoptic stylet scope (FSS) that has a simple form of a standard stylet with the fiberoptic view and maneuverability of its tip. This study was undertaken to prospectively evaluate the effectiveness of the FSS in tracheal intubation.
Methods:
Thirty-two patients undergoing general surgery participated in this study. The authors used a standard laryngoscope only to elevate the tongue, then tracheal intubation was attempted with the glottic opening being viewed only through the FSS. The success rate, time necessary for intubation, hemodynamics, and adverse effects were recorded.
Results:
The success rate of tracheal intubation on the first attempt using the FSS was 94% (30 of 32 patients), and the remaining two patients were intubated successfully on the second attempt. The mean time necessary for the intubation procedure was 29+/-14 s in all patients (mean +/- SD). Changes in hemodynamics during intubation were well within acceptable ranges. There were no major adverse effects, but minor sore throat (28%) and minor hoarseness (25%) on the first postoperative day.
Conclusions:
Tracheal intubation using the FSS proved to be a simple and effective technique for airway management.
Related Concept Videos
Cardiopulmonary Resuscitation V: Advanced Airway Management Techniques
Endotracheal Intubation I: Procedure
The ET tube comprises various components, including a standard adaptor to attach a bag-valve-mask (BVM) or ventilator, a cuff, a pilot balloon, and radiopaque markings along its length to measure the insertion distance. The tube sizes...
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Endotracheal Tube Extubation
Procedure
Extubation removes the endotracheal tube (ETT) from the patient on mechanical ventilation. It requires a well-coordinated, multidisciplinary approach involving physicians, nurses, respiratory therapists, and other healthcare professionals.
Tracheostomy Suctioning II: Procedure
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...