Alteration in flow delivery with antisyphon devices.
C McCarroll1, D McAtamney, R Taylor
1Royal Belfast Hospital for Sick Children, 180 Falls Road, Belfast BT12 6BE, UK.
Anaesthesia
|April 26, 2000
Summary
Antisyphon devices in syringe driver infusions significantly delay initial flow, especially at lower rates. Clinicians should be aware of this start-up time to ensure timely medication delivery.
Area of Science:
- Medical Devices
- Pharmacology
- Clinical Engineering
Background:
- Syringe drivers are commonly used for medication delivery.
- Antisyphon devices are incorporated into some infusion sets to prevent backflow.
- The impact of these devices on infusion start-up time is not well-established.
Purpose of the Study:
- To evaluate the effect of antisyphon devices on the time to initial flow from syringe drivers.
- To compare the start-up time of infusion sets with different antisyphon devices against a control.
Main Methods:
- Tested antisyphon devices from B Braun, Wescott, and Vygon.
- Integrated devices between a 50-ml syringe and extension set, primed with saline.
- Measured start-up time in four identical syringe drivers at 2, 10, and 50 ml/h, repeated 15 times each.
Main Results:
- At 2 ml/h, all antisyphon devices significantly increased start-up time compared to the control (p < 0.0001).
- The delay in initial flow was less pronounced at higher infusion rates (10 and 50 ml/h).
- Variability in start-up times was observed across different antisyphon device manufacturers.
Conclusions:
- Antisyphon devices can introduce a clinically relevant delay in the onset of infusion from syringe drivers, particularly at low flow rates.
- Healthcare professionals must recognize this delay to prevent potential under-infusion or delayed therapy.
- Further investigation into optimizing antisyphon device design or alternative strategies to mitigate start-up delays is warranted.