Related Experiment Videos
Recent developments in ultra low-level gamma-ray spectrometry at IRMM
Hult1, Martinez Canet MJ, Kohler
1EC-JRC-IRMM, Geel, Belgium. hult@irmm.jrc.be
Summary
The Institute for Reference Materials and Measurements (IRMM) enhanced its ultra low-level gamma-ray spectrometry capabilities with a new detector. This upgrade improves background characteristics for nuclear safeguards and material analysis.
Area of Science:
- Nuclear physics and instrumentation
- Low-background radiation detection
- Applied metrology
Background:
- Ultra low-level gamma-ray spectrometry is crucial for sensitive measurements.
- The underground HADES laboratory offers a unique low-background environment.
- Previous detectors at IRMM had limitations in background characteristics.
Purpose of the Study:
- To introduce and characterize a new low-background coaxial HPGe detector.
- To evaluate the performance of the upgraded shielding for gamma-ray spectrometry.
- To demonstrate applications in nuclear safeguards and material selection.
Main Methods:
- Installation and operation of a third coaxial HPGe detector with 60% relative efficiency.
- Construction of a new shield using 14 cm of lead (210Pb specific activity 20 Bq kg-1) and 14 cm of electrolytic copper.
- Measurement of background count rate in the HADES underground laboratory.
Main Results:
- The new detector exhibits improved background characteristics compared to previous systems.
- The integrated background count rate from 40 to 2700 keV is 394 ± 2 counts per day.
- Detailed background spectra and shielding performance are presented.
Conclusions:
- The upgraded facility significantly enhances ultra low-level gamma-ray spectrometry capabilities.
- The new detector and shielding are suitable for sensitive applications in nuclear safeguards.
- The system facilitates precise material selection and characterization through advanced spectrometry.