Related Experiment Video
Updated: Aug 19, 2026

Natural Product Discovery with LC-MS/MS Diagnostic Fragmentation Filtering: Application for Microcystin Analysis
Published on: May 31, 2019
Detection of microcystins using electrospray ionization high-field asymmetric waveform ion mobility mass
1Department of Public Health Sciences, University of Alberta, Edmonton, AB, Canada T6G 2G3.
Abstract:
A combination of electrospray ionization, high-field asymmetric waveform ion mobility spectrometry, and mass spectrometry (ESI-FAIMS/MS) was used to analyze standard solutions of microcystins-LR, -RR, and -YR. The ability of FAIMS to separate ions in the gas phase reduced the amount of background in the mass spectrum without compromising the absolute signal for these microcystins. This reduction in background resulted in a ten-fold improvement in the signal-to-background ratio over conventional ESI-MS. Detection limits, using direct infusion, were determined to be 4, 2, and 1 nM for microcystins-LR, -RR, and -YR, respectively.
More Related Videos
Related Concept Videos
Electrospray Ionization (ESI) Mass Spectrometry
ESI utilizes electrical energy to transfer ions from the liquid phase of the sample into the...
MALDI-TOF Mass Spectrometry
Peptide Identification Using Tandem Mass Spectrometry
This technique helps gather information regarding the protein from which the peptide was obtained and to study the peptides’ amino acid sequence. Identifying peptides from a complex mixture is an important component of the growing field of...
Mass Spectrometry: Complex Analysis
GC–MS is a powerful hyphenated method commonly used in forensics and environmental...
Tandem Mass Spectrometry
High-Performance Liquid Chromatography: Types of Detectors