Related Experiment Videos
Structured sleeve for repair of implantable in-line connectors
P Strojnik1, T G Stage, N Bhadra
1Department of Biomedical Engineering, Case Western Reserve University, Cleveland, Ohio, USA. prs3@po.cwru.edu
Medical & Biological Engineering & Computing
|September 14, 2000
Abstract:
A structured miniature repair sleeve has been designed for implantable in-line connectors that develop small current leaks post-implant. The repair sleeve has been successfully utilised in one subject following the development of current leakage in connectors on an implanted joint angle sensor.