Feasibility and initial results of an Internet-based pacemaker and ICD pulse generator and lead registry

R Hauser1, D Hayes, V Parsonnet

  • 1Minneapolis Heart Institute Foundation, 920 E. 28th St., Ste #300, Minneapolis, MN 55407, USA. rhauser747@aol.com

Insights

An online registry for pacemaker and implantable cardioverter-defibrillator (ICD) device failures provides timely data on performance issues. This system identified new problems and offered insights into device longevity and failure causes.

Area of Science:

  • Cardiology
  • Biomedical Engineering
  • Medical Device Technology

Background:

  • Lack of independent, timely data on pacemaker and implantable cardioverter-defibrillator (ICD) device performance.
  • Need for a system to track pulse generator and lead failures in real-time.

Purpose of the Study:

  • To establish and evaluate an Internet-based registry for tracking pacemaker and ICD device failures.
  • To provide timely information on device performance, failure modes, and clinical consequences.

Main Methods:

  • Development and implementation of an online registry (www.pacerandicregistry.com) for reporting device failures.
  • Analysis of data collected during the first year of registry operation.

Main Results:

  • Identification of three previously unreported device problems, communicated promptly to participants.
  • 11% of ICD and 10% of pacemaker pulse generator failures occurred before the elective replacement indicator (ERI).
  • Average battery longevity: 4.0 years for ICDs, 6.8 years for dual-chamber pacemakers. Lead failures: 54% pacemaker and 29% ICD due to disrupted insulation. ICD failures had more severe clinical consequences.

Conclusions:

  • An Internet-based registry is a feasible method for collecting and disseminating timely data on pacemaker and ICD failures.
  • The registry provides valuable insights into device failure signs, causes, and clinical impact.
  • This system can enhance patient safety and inform medical practice regarding device performance.