Characterization of formaldehyde by gas chromatography using multiple pulsed-discharge photoionization detectors and a flame ionization detector
J G Dojahn1, W E Wentworth, S D Stearns
1University of Houston, Department of Chemistry, TX 77024-5641, USA.
Abstract:
Formaldehyde, water, methanol, butanal, and butanone are characterized by gas chromatography using three pulsed-discharge photoionization detectors (PDPIDs) and a flame ionization detector (FID). One of the PDPIDs is operated in helium mode, and the other two are operated in argon and krypton modes. The FID is included for comparison. The PDPIDs are used to efficiently differentiate between and quantitatively identify formaldehyde and the other three compounds in a sample mixture. This is accomplished by using butanone as the internal standard and correlating the relative responses of the four organic compounds in the helium-, argon-, and krypton-mode PDPIDs with their relative retention times.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
Flame Photometry: Lab
Flame Photometry: Overview
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...