Determining the accuracy of articulator interchangeability and hinge axis reproducibility
1School of Dental Medicine, Tufts University, Boston, Mass, USA.
The Journal of Prosthetic Dentistry
|March 27, 2001
Summary
Dental articulators are not exact duplicates, but the Artex brand offers the most consistent hinge axis repeatability and interchangeability for accurate dental cast remounting.
Area of Science:
- Dental Technology
- Prosthodontics
- Occlusion
Background:
- Articulators are assumed to be interchangeable for accurate dental laboratory remounting of casts.
- Clinicians need assurance of positional accuracy when transferring casts between articulators.
Purpose of the Study:
- To evaluate the positional accuracy of transferred dental casts between articulators.
- To determine the hinge axis reproducibility of tested articulator systems.
Main Methods:
- Compared occlusal contact areas of mounted casts transferred between like articulators (Whip Mix 3040, Girrbach Artex AL, KaVo Protar).
- Utilized vinyl polysiloxane interocclusal records and computerized image analysis for quantitative measurements.
- Assessed positional changes using a calibrated grid and Kruskal-Wallis statistical tests.
Main Results:
- No articulator system demonstrated exact duplication or interchangeability (P<.01).
- Whip Mix articulators showed greater deviation in hinge axis repeatability and interchangeability compared to KaVo.
- Girrbach Artex AL articulators exhibited the most consistent hinge axis repeatability and interchangeability.
Conclusions:
- The Artex brand provided superior consistency in reproducing dental cast positions compared to other tested articulators.
- Significant deviations exist between articulator systems, impacting positional accuracy.