Related Experiment Video
Updated: Aug 17, 2026

In vitro Assessment of Aortic Regurgitation Using Four-Dimensional Flow Magnetic Resonance Imaging
Published on: February 25, 2022
Flow dependency of error in thermodilution measurement of cardiac output during acute tricuspid regurgitation
P M Heerdt1, G A Blessios, M L Beach
1Department of Anesthesiology, New York Presbyterian Hospital-Weill Medical College at Cornell University, New York, NY, USA.
Objective:
To evaluate the effect of variable degrees of tricuspid regurgitation on thermodilution cardiac output measurements during changes in venous return.
Design:
Prospective, controlled animal study.
Setting:
University laboratory.
Participants:
Eight anesthetized mongrel dogs instrumented for simultaneous measurement of cardiac output by thermodilution and ascending aortic electromagnetometry.
Interventions:
Data were collected before and after induction of moderate and severe tricuspid regurgitation. Under each condition, measurements were obtained at baseline and after opening 2 peripheral arteriovenous shunts to increase venous return.
Measurements And Main Results:
Baseline electromagnetic flow ranged from 1.74 to 3.62 L/min (median 2.73 L/min). Moderate and severe regurgitation reduced median electromagnetic flow values by 31% and 51%. Applying generalized estimating equations to model thermodilution cardiac output as a function of electromagnetic flow, arteriovenous shunt, and severity of tricuspid regurgitation revealed that (1) moderate and severe regurgitation changed the slope and intercept of the thermodilution/electromagnetic regression, but the differences between them were not significant, and (2) arteriovenous shunt alone had no effect under any condition. A simplified model independent of shunt and containing just 2 levels of tricuspid regurgitation (none or present) crossed with electromagnetic flow was applied. This analysis showed that regurgitation caused thermodilution to significantly underestimate electromagnetic flow at cardiac outputs > 2.27 L/min (99 mL/kg/min) and overestimate it at flows < 1.02 L/min (44 mL/kg/min).
Conclusions:
These data show that acute tricuspid regurgitation may produce underestimation of cardiac output by thermodilution when flow is relatively high, produce overestimation when flow is relatively low, or have minimal effect when flow is in the midrange.
Related Concept Videos
Imbalances in Cardiac Output
CHF can occur due to the failure of either side of the heart. Left-side failure leads to pulmonary congestion—the right side continues to send blood...
Cardiac Output I:Effect of Heart Rate on Cardiac Output
Cardiac output (CO) refers to the total amount of blood ejected by one of the ventricles in liters per minute (L/min). In a resting adult, CO ranges from 5 to 6 L/min, adjusting according to the body's metabolic requirements.
Effect of Heart Rate on Cardiac Output
Cardiac output adapts to metabolic demands during stress, physical activity, or illness. The autonomic nervous system regulates heart rate via the sinoatrial node. The parasympathetic nervous system decreases heart rate...
Cardiac Output II: Effect of Stroke Volume on Cardiac Output
Preload
Preload refers to the initial elongation of the cardiac myocytes before contraction and is related to the volume of blood filling the heart at the end of diastole, or end-diastolic volume. The...
Mitral Regurgitation I: Introduction
Aortic Regurgitation I: Introduction
Aortic Regurgitation II: Clinical Features and Diagnostic Tests