E T Huang1, K R Hardy, J M Stubbs
1Institute for Environmental Medicine, Department of Emergency Medicine, Department of Epidemiology and Biostatistics, Center for Epidemiology and Biostatistics, University of Pennsylvania School of Medicine, Philadelphia, Pennsylvania, USA.
Abstract:
A novice scuba diver with an implanted ventriculo-peritoneal (VP) shunt inquired about the performance characteristics of his shunt while diving. A literature search revealed no information regarding shunt performance under hyperbaric conditions. The manufacturer could not certify that the shunt would function under pressure. Therefore, four VP shunts were tested according to the manufacturer's testing protocol at 1 and 4 atm abs in a multiplace hyperbaric chamber. The pressure (in mm of H2O) required to establish flow through the shunts was recorded. Trials at 1 atm abs (n = 12) and 4 atm abs (n = 12) show that all shunts performed within the pressure range specified by the manufacturer.