A comparison of quantitative coronary angiography systems using a unique set of in vivo coronary stenosis images
G B Mancini1, E Yeoh, C Kamimura
1University of British Columbia, Vancouver Hospital and Health Sciences Centre, Vancouver, Canada. deptmed@vanhosp.bc.ca
Objective:
To compare the accuracy and precision of measurements of minimum lumen diameter (MLD) among two existing, and one new, quantitative coronary angiography systems.
Materials And Methods:
The analysis was performed using in vivo cinearteriograms of precisely drilled, radiolucent plastic beads that were inserted percutaneously into the coronary arteries of canines. The existing algorithms compared were the ArTrek and the Coronary Measurement System (CMS). The latter was applied in two modes: a mode based on a minimal cost analysis algorithm and a mode based on a gradient field transform. The new algorithm (CorTrek) was also applied in two modes: a mode called the ArTrek compatible mode and a mode known as the 'regression' mode. The latter mode uses a look-up table, based on a phantom calibration step, to readjust the measured MLD to overcome system nonlinearities (overestimation of small diameters and underestimation of large diameters).
Results:
In the absence of editing, the optimal accuracy (no significant bias) was achieved with the ArTrek compatible algorithm (-0.03 mm for an MLD between 0.83 and 1.83 mm). All other algorithms showed significant under- or overestimation of the MLD within this range. The precision ranged from 0.18 to 0.40 mm without editing, and was best with the ArTrek algorithm.
Conclusions:
The present study provides in vivo validation and comparative performance characteristics of a new, accurate coronary quantitative angiography system.
More Related Videos
06:39Ultrasound Based Assessment of Coronary Artery Flow and Coronary Flow Reserve Using the Pressure Overload Model in Mice
Published on: April 13, 2015
06:18Intravascular Ultrasound Image-Based Finite Element Modeling Approach for Quantifying In Vivo Mechanical Properties of Human Coronary Artery
Published on: December 6, 2024
Related Concept Videos
Imaging Studies for Cardiovascular System V: CT
Imaging Studies for Cardiovascular System VI: Calcium -Scoring CT
Acute Coronary Syndrome III: Diagnostic Studies