Related Experiment Video
Updated: Oct 5, 2026

Unraveling the Unseen Players in the Ocean - A Field Guide to Water Chemistry and Marine Microbiology
Published on: November 5, 2014
Estimation of dissolved organic carbon (DOC) concentrations in nanoliter samples using UV spectroscopy
1Institut des Sciences de la Mer de Rimouski, Université du Québec à Rimouski, Canada.
Abstract:
A simple analytical system has been developed and tested for estimating dissolved organic carbon (DOC) concentrations in nanoliter samples of sediment porewaters. The system consists of a loop injector that introduces 200 nL of sample directly into the capillary tubing connected to a scanning UV-VIS detector equipped with a 35-nanoliter capillary flowcell. Analysis of DOC by spectroscopic and high-temperature catalytic oxidation (HTCO) methods on the same samples showed a strong linear correlation between UV absorbance at 254 nm and HTCO-DOC concentration, allowing the DOC quantification. The simplicity and the robustness of the UV system allow DOC measurements in the field at a rate of 30 samples/h. We have used the UV system successfully for determining high-resolution distributions of porewater DOC concentrations in sediments at millimeter scales.
Related Concept Videos
UV–Vis Spectrometers
Ultraviolet and Visible (UV–Vis) Spectroscopy: Overview
Atomic Absorption Spectroscopy: Lab
Solutions containing organic solvents, such as low-molecular-mass alcohols, esters, or ketones, enhance absorbances by increasing nebulizer...
UV–Vis Spectroscopy: Beer–Lambert Law