Absorbed doses from spiral CT and conventional spiral tomography: a phantom vs. cadaver study
C Bou Serhal1, R Jacobs, F Gijbels
1Department of Periodontology, Oral Imaging Cluster, School of Dentistry, Oral Pathology and Maxillofacial Surgery, Catholic University Leuven, Kapucijnenvoer 7, 3000 Leuven, Belgium.
Abstract:
For several radiological examinations, a clinician can select between conventional and spiral computed tomography. Using both techniques, this study aimed at evaluating the difference in absorbed doses when examining a single lateral jaw segment in a human cadaver head and Rando phantom. The present study involved the placement of thermoluminescent dosimeter (TLD) chips (GR-200) in the thyroid gland, and bilaterally, in the parotid and submandibular glands and the lenses of the eyes in both a human cadaver and a Rando phantom at corresponding locations. Consecutive conventional spiral tomographic examinations were carried out in both the left upper and lower premolar area, using a Cranex TOME multifunctional unit. Each examination consisted of 4 slices with a 2 mm slice thickness and exposure parameters of 57 kV, 56 seconds and 1.6-2.0 mA. Regarding spiral computed tomography (CT), a Somatom Plus S scanner (Siemens, Erlangen, Germany), with a slice thickness of 1 mm with settings at 120 kV and 165 mA, was used on both phantoms and separately in the upper and lower jaw. With conventional tomography, the findings of the present study showed that the parotid and submandibular glands on the side near the X-ray tube received the highest dose, both for the cadaver head (doses ranging from 0.5 to 1.3 mGy) and the phantom (doses ranging from 0.6 to 2.6 mGy). For CT of the upper jaw, the highest doses were delivered to the parotid glands with an average absorbed dose of 9.2 and 10.6 mGy for the cadaver head and phantom, respectively. The submandibular glands received the highest doses during CT examination of the lower jaw with an average of 7.8 and 12.9 mGy for the cadaver head and phantom, respectively. It appears from the present investigation that if small edentulous regions are examined, radiation doses during conventional tomography remain much lower than during CT imaging. However, when multiple tomographic cuts are required, a spiral CT examination can replace a series of conventional examinations, especially in cases such as the rehabilitation of an edentulous upper jaw or a more complex surgery.
More Related Videos
08:41Patient-Specific Polyvinyl Alcohol Phantom Fabrication with Ultrasound and X-Ray Contrast for Brain Tumor Surgery Planning
Published on: July 14, 2020
12:32Image Rendering Techniques in Postmortem Computed Tomography: Evaluation of Biological Health and Profile in Stranded Cetaceans
Published on: September 27, 2020
Related Concept Videos
Computed Tomography
The technique was invented in the 1970s and is based on the principle that as X-rays pass through the body, they are absorbed or reflected at different levels. In the technique, a patient lies on a motorized platform while a computerized axial tomography (CAT) scanner rotates...
Imaging Studies I: CT and MRI
Description of the Procedures
Computed Tomography (CT) scan:
Computed Tomography (CT) scans use X-ray technology to generate detailed images of bones, organs, and tissues. During the scan, the patient lies on a moving table...
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
Imaging Studies for Cardiovascular System V: CT
Imaging Studies III: Computed Tomography