Observation of sorptive losses of volatile sulfur compounds during natural gas sampling
Michael Sulyok1, Christina Haberhauer-Troyer, Erwin Rosenberg
1Institute of Analytical Chemistry, Vienna University of Technology, Austria.
Abstract:
The reason for the compound-dependent over-estimation of the recoveries of several volatile organic sulfur compounds when using a Silcosteel cylinder for sample storage as reported earlier was examined. From the different possible sources of errors that were taken into consideration, the silicone tubing, which was used to fill a standard Tedlar sample bag for calibration, was identified as the cause of the artefact. The comparison of different tubing materials showed that PTFE is the best choice since it causes only minor losses (<10%) of propyl- and butylmercaptans.
Related Concept Videos
Volatilization
Sampling Methods: Sample Types
Solid samples include a variety of substances, such as sediments from water bodies, soil, metals, and biological tissues. Two standard methods for extracting sediments from water bodies are grab sampling and piston coring. Grab sampling involves using a device to collect a discrete sediment sample from the bottom of a water body with minimal disturbance. Grab samples do not always represent the entire area due to...