Dosimetric characteristics with spatial fractionation using electron grid therapy.

A S Meigooni1, S A Parker, J Zheng

  • 1University of Kentucky, Chandler Medical Center, Department of Radiation Medicine, Lexington 40536, USA. alimeig@pop.uky.edu

Medical Dosimetry : Official Journal of the American Association of Medical Dosimetrists
|May 22, 2002
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Quantitative verification of <sup>192</sup> Ir PDR and HDR source structure by pin-hole autoradiography.

Medical physics·2018
Same author

SU-E-T-320: A New Verification Phantom for GYN Brachytherapy Applicators Using GafChromic - Films.

Medical physics·2017
Same author

A New Approach for Heterogeneity Corrections for Cs-137 Brachytherapy Sources.

Journal of biomedical physics & engineering·2015
Same author

EchoSeed Model 6733 Iodine-125 brachytherapy source: improved dosimetric characterization using the MCNP5 Monte Carlo code.

Medical physics·2012
Same author

A novel device for automatic withdrawal and accurate calibration of 99m-technetium radiopharmaceuticals to minimise radiation exposure to nuclear medicine staff and patient.

Radiation protection dosimetry·2012
Same author

Dosimetric characteristics of the new RadioCoil 103Pd wire line source for use in permanent brachytherapy implants.

Medical physics·2004

Spatially fractionated electron grid therapy shows promise for treating bulky superficial lesions. A 2.5-cm hole diameter in cerrobend blocks optimized dose distributions for electron beams (6-20 MeV), offering a new palliative treatment option.

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Radiotherapy

Background:

  • Megavoltage photon grid therapy shows clinical promise for palliative treatments.
  • Its application is limited for bulky superficial lesions near critical structures.
  • Spatially fractionated electron grid therapy is explored as an alternative.

Purpose of the Study:

  • To investigate the dose distributions of electron beams using cerrobend grid blocks.
  • To determine the optimal hole diameter for spatially fractionated electron grid therapy.
  • To evaluate the feasibility of this technique for superficial lesion treatment.

Main Methods:

  • Experimental determination of dose distributions using 1.4-cm-thick cerrobend grid blocks.
  • Electron beams ranging from 6 to 20 MeV were used.

Related Experiment Videos

  • Measurements of beam profiles and percentage depth dose (PDD) curves were performed using film, TLD, and ionometry in Solid Water phantom.
  • Comparison of PDD curves with various grid hole diameters (1.5-3.5 cm) to identify the optimal size.
  • Main Results:

    • A 2.5-cm hole diameter was identified as optimal for all tested electron energies (6-20 MeV).
    • Peak-to-valley ratios decreased with depth, with the largest ratio observed at the dose maximum (Dmax).
    • Dose under blocked regions ranged from 9.7% to 39% of the dose under grid holes, varying with location and energy.

    Conclusions:

    • Spatially fractionated electron grid therapy is a viable technique for delivering palliative radiation.
    • The 2.5-cm hole diameter provides optimal dose fractionation for electron beams in this setup.
    • This method offers potential for treating superficial lesions while sparing surrounding healthy tissues.