Protocol for Relative Hydrodynamic Assessment of Tri-leaflet Polymer Valves
Published on: October 17, 2013
Experimental comparison between a Staffieri and a New Prototype tracheo-oesophageal valve
Guido Belforte1, Massimiliana Carello
1Department of Mechanics, Technical University of Torino, C. so Duca degli Abruzzi 24, 10129 Torino, Italy.
Abstract:
A tracheo-oesophageal valve is used for the rehabilitation of patients who have lost vocal function due to a total laryngectomy. The valve is positioned in a surgically created fistula between the airway and the upper oesophagus at the level of the post-laryngectomy stoma. It permits passage of air from the trachea to the pharynx to permit speech, while it blocks the flow of secretions and food materials from the digestive tract to the airway. In this paper, experimental results obtained with a Staffieri and a New Prototype tracheo-oesophageal valve are presented. The valves have been tested experimentally under different conditions of airflow through the valve and tracheal-side pressure; the airflow resistance has been calculated to compare the performance of the valves. The valves have also been experimentally tested under different conditions of fluid flow through the valve and oesophageal-side pressure. The airflow resistance of the Staffieri and the New Prototype valves has been compared for different angular extents, alpha, of the razor-thin slit through which the airflow passes. For each type of valve the parameter alpha has a significant influence on flow characteristics, while the shape of the dome becomes important only for the larger angular extents considered.Regarding the reverse flow, it is pointed out that for both of the valve types at different oesophageal pressures and for smaller angular extents of the razor-thin slit, the reverse fluid flow through the valves is smaller than the allowable reverse flow, that can be tolerated by patients without creating problems. In general, for high values of the angular extent of the razor-thin slit the Staffieri valve offers the least flow resistance, but, unlike the New Prototype, the direct flow under these conditions in unacceptably large.
Related Concept Videos
Tracheostomy: Procedure and Tubes
Tracheostomy tubes can be made of semiflexible plastic (polyurethane or silicone), rigid plastic, or metal, and they come in...
Oxygen Delivering System III: Tracheostomy and T-piece
Tracheostomy
A tracheostomy is a surgically created opening (stoma) in the anterior part of the trachea. It is used to establish a patient airway, bypass an upper airway obstruction, simplify the removal of secretions, permit long-term...
Oxygen Delivering System II: Venturi Mask and Transtracheal Oxygen
Venturi Mask
The Venturi mask, named after the Venturi effect, is designed to deliver precise oxygen concentrations. It consists of a large tube with an oxygen inlet that narrows down, causing a pressure drop that pulls air in through adjustable side ports. The mask is a lightweight,...
Tracheostomy Care II: Procedure
Step 1: Perform hand hygiene, and put on personal protective equipment: gown, gloves, mask and...
Tracheostomy Suctioning II: Procedure
Tracheostomy Decannulation
Description of the Procedure
Decannulation refers to the permanent removal of the tracheostomy tube, signaling the resolution of the condition that initially necessitated the tracheostomy. The process requires a well-coordinated interplay between...