Summary
Percutaneous suture closure using the Perclose Techstar XL device effectively achieves hemostasis at brachial artery puncture sites in most patients, including those on anticoagulation. Obesity may be a factor in successful application for brachial artery closure.
Area of Science:
- Vascular Surgery
- Interventional Cardiology
- Medical Device Technology
Background:
- Percutaneous suture closure offers immediate hemostasis for arterial access sites.
- The Perclose Techstar XL device is a novel tool for achieving hemostasis.
- Brachial artery access is utilized for various diagnostic and therapeutic catheterizations.
Purpose of the Study:
- To evaluate the efficacy and safety of the Perclose Techstar XL 6Fr suture closure device for brachial artery hemostasis.
- To assess the applicability of this device in managing brachial artery puncture sites post-catheterization.
Main Methods:
- Ten procedures were performed on eight patients using the Perclose Techstar XL 6Fr device for brachial artery hemostasis.
- Both 6Fr and 7Fr sheaths were employed during the procedures.
- Hemostasis was monitored post-catheterization.
Main Results:
- Successful hemostasis was achieved in nine out of ten procedures (90%).
- One patient required surgical repair due to failed hemostasis.
- The device demonstrated utility in selected patients for brachial artery puncture site management.
Conclusions:
- The Perclose Techstar XL device is a viable option for brachial artery hemostasis in select patients.
- The device's large foot excursion may limit its applicability in smaller arteries like the brachial artery.
- Obese patients may be particularly well-suited for this percutaneous closure method.