Errors in repolarization measurement using magnetocardiography.

Fiona E Smith1, Philip Langley, Lutz Trahms

  • 1Regional Medical Physics Department, Freeman Hospital, Newcastle-upon-Tyne, UK,. Fiona.Smith@nuth.northy.nhs.uk

Pacing and Clinical Electrophysiology : PACE
|October 3, 2002
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Muscle energetics and the cardiovascular response to isometric exercise and post-exercise circulatory occlusion following exercise-induced muscle damage: insights from multiparametric MRI.

Frontiers in physiology·2025
Same author

Activity-Based Costing in IVF: a framework for transparency and operational scaling of fertility services.

Journal of assisted reproduction and genetics·2025
Same author

Carbohydrate intake of 10 g/kg body mass rapidly replenishes liver, but not muscle glycogen contents, during 12 h of post-exercise recovery in well-trained cyclists.

The Journal of physiology·2025
Same author

A research agenda for GIScience in a time of disruptions.

International journal of geographical information science : IJGIS·2025
Same author

A digitally controlled, remotely operated ICSI system: case report of the first live birth.

Reproductive biomedicine online·2025
Same author

Preventing Cardiomyopathy in Duchenne Muscular Dystrophy: Long-Term Follow-Up of Patients in the Randomised, Placebo-Controlled Drug-Trial of Perindopril and Bisoprolol.

European journal of neurology·2025
Same journal

Vein of Marshall Alcohol Ablation Facilitated by a 10-French Bidirectional Catheter Supported by Glide Advantage Wire.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Leadless Pacemaker Implantation in High-Risk Patients: Real-World Outcomes From a Multicenter Israeli Registry.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Sex Differences in the Burden of Atrial Fibrillation/Flutter and Associated Heart Failure Stratified by Age at Onset.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Syncope due to Pacing Inhibition During VT Detection in a Conduction System Pacing ICD Programmed in AAI(R) Mode.

Pacing and clinical electrophysiology : PACE·2026
Same journal

When Less Is Too Little: Low Atrioventricular Synchrony by VDD Leadless Pacemaker in a Patient With Glenn Circulation.

Pacing and clinical electrophysiology : PACE·2026
Same journal

Temporal Changes in Left Ventricular Strain in Typical and Atypical Left Bundle Branch Block Patients: A Follow-up Study.

Pacing and clinical electrophysiology : PACE·2026
See all related articles

Manual measurement of repolarization intervals using multichannel magnetocardiography (MCG) is affected by display settings and noise. Standardized presentations are crucial for accurate MCG repolarization measurements.

Area of Science:

  • Biophysics
  • Cardiology
  • Medical Imaging

Background:

  • Multichannel magnetocardiography (MCG) noninvasively measures magnetic fields from the heart.
  • MCG offers potential for regional ventricular repolarization assessment.
  • Quantified errors exist for electrocardiography (ECG) repolarization, but not for MCG.

Purpose of the Study:

  • To determine errors in manual repolarization interval measurements in MCG.
  • To evaluate the impact of display characteristics and noise on MCG measurements.
  • To compare inter-analyst variability in MCG repolarization interval measurements.

Main Methods:

  • Simultaneous recording of 16-channel MCG and 3-lead ECG in 8 healthy subjects.
  • Manual measurement of repolarization intervals across 2,048 MCG presentations with varied noise, time, and amplitude displays.

Related Experiment Videos

  • Four analysts performed measurements on randomized recordings.
  • Main Results:

    • Added noise reduced intervals by 3-6 ms.
    • Increased time display width shortened intervals by 9-19 ms.
    • Increased amplitude display height increased measurements by 7 ms.
    • Inter-analyst differences reached a maximum mean of 24 ms.

    Conclusions:

    • Display characteristics, added noise, and analyst variability significantly impact manual MCG repolarization interval measurements.
    • Standardized presentation protocols are essential for reliable MCG repolarization assessment.
    • Further research is needed to establish robust MCG measurement standards.