Usefulness of laboratory neutron fields
1Atomic Energy of Canada Ltd. (AECL), Chalk River Laboratories, Chalk River, Ontario, Canada K0J 1J0. nunesj@aecl.ca
Radiation Protection Dosimetry
|April 8, 2003
Summary
This study tested a neutron survey meter for nuclear power plants, comparing measured and theoretical calibration constants across various neutron fields. Results offer insights into potential over- and under-response in workplace settings.
Area of Science:
- Nuclear Engineering
- Radiation Detection and Measurement
Background:
- Neutron survey meters are crucial for radiation protection in nuclear power plants.
- Accurate calibration is essential for reliable dose assessment in diverse neutron fields.
Purpose of the Study:
- To evaluate the performance of a moderator-type neutron survey meter.
- To compare measured and theoretically calculated calibration constants.
- To assess the meter's response in simulated reactor workplace neutron fields.
Main Methods:
- Testing the neutron survey meter in five laboratory neutron fields.
- Determining calibration constants and responses by measurement.
- Calculating theoretical calibration constants using a literature response function.
- Analyzing meter response against reference ambient dose equivalent values.
Main Results:
- Measured and theoretical calibration constants showed good agreement.
- The study identified the range of potential over- and under-response for the survey meter.
- Calibration constants were determined for various neutron spectra, including simulated power-plant workplaces.
Conclusions:
- The tested neutron survey meter demonstrates predictable performance across different neutron fields.
- Theoretical calculations can reliably estimate calibration constants for this type of survey meter.
- Findings provide guidance for using the survey meter effectively in nuclear power plant environments.