Possible application of an imaging plate to space radiation dosimetry
Related Experiment Video
Updated: Sep 25, 2026

Autoradiography as a Simple and Powerful Method for Visualization and Characterization of Pharmacological Targets
Published on: March 12, 2019
Hiroko Ohuchi1, Akira Yamadera
1Graduate School of Pharmaceutical Sciences, Tohoku University, Aramaki-Aoba, Aoba-ku, Sendai 980-8578, Japan. hiroko@mail.pharm.tohoku.ac.jp
Abstract:
Fading correction plays an important role in the application of commercially available BaBrF:Eu2+ phosphors: imaging plates (IP) to dosimetry. We successfully determined a fading correction equation, which is a function of elapsed time and absolute temperature, as the sum of several exponentially decaying components having different half-lives. In this work, a new method was developed to eliminate a short half-life component by annealing the IP and estimating the radiation dose with the long half-life components. Annealing decreases the effect of fading on the estimated dose, however, it also causes the loss of photo-stimulated luminescence (PSL). Considering an IP as an integral detector for a specific period of up to one month, the practically optimum conditions for quantitative measurement with two types of IP (BAS-TR and BAS-MS) were evaluated by using the fading correction equation, which was obtained after irradiation with a 244Cm source as the alpha-ray source having a specific radioactivity of 1,638.5 Bq/cm2 including beta and gamma-ray (alpha energy of 5.763 and 5.805 MeV). Annealing at 80 degrees C for 24 hours after irradiation for one month using BAS-MS should minimize the effect of the elapsed time, resulting in sufficient sensitivity. The results demonstrate new possibilities for radiation dosimetry offered by the use of an IP.
More Related Videos
06:20Irradiator Commissioning and Dosimetry for Assessment of LQ α and β Parameters, Radiation Dosing Schema, and in vivo Dose Deposition
Published on: March 11, 2021
06:28Visualization of Low-Level Gamma Radiation Sources Using a Low-Cost, High-Sensitivity, Omnidirectional Compton Camera
Published on: January 30, 2020
Related Concept Videos
Imaging Studies II: Positron Emission Tomography and Scintigraphy
Fundamental Principles of PET
X-ray Imaging
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...
Determination of Crystal Structures
Biological Effects of Radiation