Compensator quality control with an amorphous silicon EPID.

Geetha V Menon1, Ron S Sloboda

  • 1Department of Physics, University of Alberta, Edmonton, Alberta, T6G 2E1, Canada.

Medical Physics
|August 9, 2003
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

Correction: "Generation and comparison of 3D dosimetric reference datasets for COMS eye plaque brachytherapy using model-based dose calculations" https://doi.org/10.1002/mp.16721.

Medical physics·2024
Same author

Prostate size, source configuration, and dosimetry dynamics of stranded <sup>125</sup>I seed implants.

Brachytherapy·2024
Same author

Generation and comparison of 3D dosimetric reference datasets for COMS eye plaque brachytherapy using model-based dose calculations.

Medical physics·2023
Same author

AAPM WGDCAB Report 372: A joint AAPM, ESTRO, ABG, and ABS report on commissioning of model-based dose calculation algorithms in brachytherapy.

Medical physics·2023
Same author

Delivered dose changes in COMS plaque-based ocular brachytherapy arising from vitrectomy with silicone oil replacement.

Brachytherapy·2019
Same author

Robotic-Assisted Needle Steering Around Anatomical Obstacles Using Notched Steerable Needles.

IEEE journal of biomedical and health informatics·2018

Amorphous silicon electronic portal imaging devices (a-Si EPID) offer a viable alternative for quality control of radiation therapy compensators. This study demonstrates a-Si EPID can accurately measure compensator factors and fluence ratios, matching ion chamber results within 2%.

Area of Science:

  • Medical Physics
  • Radiation Oncology
  • Imaging Technology

Background:

  • Compensator calibration and quality control are crucial in radiation therapy.
  • Conventional methods use ion chambers in water-equivalent phantoms.
  • An alternative, efficient method is needed for quality assurance.

Purpose of the Study:

  • To investigate the suitability of amorphous silicon electronic portal imaging devices (a-Si EPID) for compensator quality control.
  • To compare a-Si EPID measurements with conventional ion chamber methods.
  • To establish a reliable and potentially faster quality control protocol.

Main Methods:

  • Determined a-Si EPID response to energy fluence for open and compensated fields.
  • Developed calibration curves accounting for linear and quadratic responses.

Related Experiment Videos

  • Measured EPID scatter factors considering field size, compensator thickness, and source-detector distance.
  • Evaluated clinical and test compensators using the a-Si EPID.
  • Main Results:

    • The a-Si EPID demonstrated a linear response for open fields and a small quadratic component for compensated fields.
    • EPID scatter factors showed minor dependencies on compensator thickness and source-detector distance.
    • Measurements of clinical compensator factors and off-axis fluence ratios were within 2% of Farmer chamber values on average.

    Conclusions:

    • The a-Si EPID is a suitable replacement for ion chambers in compensator quality control.
    • This technique offers a promising alternative for routine quality assurance in radiation therapy.
    • The findings support the integration of a-Si EPID into clinical workflows for enhanced efficiency.