Related Experiment Video
Updated: Sep 20, 2026

Rotating the Intraocular Lens to Prevent Posterior Capsular Opacification in Cataract Surgeries
Published on: July 7, 2023
Lens epithelial cell migration between posterior capsule and intraocular lens with variously finished posterior optic
Masahiko Ayaki1, Yoshiko Ishida, Eiichi Nishimura
1Department of Ophthalmology, Fujigaoka Hospital, Showa University School of Medicine, Yokohama, Japan. mayaki@olive.ocn.ne.jp
Purpose:
To compare lens epithelial cell migration between intraocular lens (IOL) optics and the posterior capsule of IOLs with various edges and haptic angulations.
Methods:
Experimental acrylic IOLs with 3 types of finishing (none, polishing, tumbling and polishing) and 2 types of haptic angulation (10 and 15 degrees ) as well as 3 (AcrySof MA30BA, Sensar AR40, CeeOn 911) clinically implanted IOLs were inserted in rabbit eyes. Scanning electron-microscopic examination of IOLs was carried out before surgery and histological examination of lens capsules 3 weeks after surgery. The grading of lens epithelial cell migration under the IOL was made by scoring.
Results:
Scanning electron-microscopic examination showed that the optic edge sharpness decreased according to the finishing in experimental IOLs. The AcrySof and CeeOn lenses had sharp edges and the Sensar had a rounded edge. The migration score was highest in the IOL with a rounded edge whereas it was lowest in IOLs with a sharp edge and large haptic angulation. There was no significant difference between the scores of clinically implanted IOLs.
Conclusion:
Sharpness of the optic edge may be a critical factor for the prevention of posterior capsule opacification and more important than haptic angulation in acrylic IOLs.
Related Concept Videos
Focusing of Light in the Eye
Accessory Structures of the Eye
Glaucoma: Overview