Optimization stacking by transient pseudo-isotachophoresis for capillary electrophoresis: example analysis of plasma
Yu Kong1, Ning Zheng, Zhichao Zhang
1Institute of Elemento-Organic Chemistry, Nankai University, 94 Weijin Road, 300071 Tianjin, China.
Abstract:
Low concentrations of both reduced form glutathione (GSH) and oxidized form glutathione (GSSG) in diabetic nephropathy (DN) patient's plasma were measured with transient pseudo-isotachophoresis. The plasma samples were deproteined with acetonitrile and centrifuged. The method was performed at constant voltage of 5 kV using a 300 mM borate buffer (pH 8.0), with a fused-silica capillary of 21 cm x 75 microm. The sample length can reach 25% of the efficient length of the capillary, and the sensitivities of GSH and GSSG increased 15-20-fold. The method was also systematically optimized, and the results show that this type of stacking offers good repeatability for routine clinical assay of glutathione in DN plasmas.
Related Concept Videos
SDS-PAGE
A variation of gel electrophoresis, termed polyacrylamide gel electrophoresis (PAGE), is commonly used for separating proteins according to their molecular size by passing them through a polyacrylamide gel. Because of the varying charges associated with amino acid side chains, PAGE can be used to separate intact proteins...
Electrophoresis: Overview
There...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...