Reliable method for the determination of ranitidine by liquid chromatography using a microvolume of plasma
Related Experiment Video
Updated: Sep 20, 2026

HPLC-based Assay to Monitor Extracellular Nucleotide/Nucleoside Metabolism in Human Chronic Lymphocytic Leukemia Cells
Published on: July 20, 2016
Carmen Flores Pérez1, Hugo Juárez Olguín, Janett Flores Pérez
1Department of Pharmacology, Instituto Nacional de Pediatría, CP 04530 Mexico City, Mexico. carmenfloresp@hotmail.com
Abstract:
The aim of the present study was to develop a simple method to measure ranitidine, using 100 microL of plasma, by high-performance liquid chromatography with a Symmetry C(18) column and UV detection at 313 nm. Linearity was assessed in the range from 50 to 1500 ng ml(-1) and had a correlation coefficient of 0.999. The inter- and intra-day coefficients of variation were less than 7%. The limits of detection and quantitation were 5 and 15 ng ml(-1), respectively. Drug levels were determined satisfactorily in three patients. A simple and reliable method was developed which uses a microvolume of plasma, particularly useful in low-weight children.
Related Concept Videos
Determination of Renal Drug Clearance: Graphical and Midpoint Methods
The graphical method involves plotting the rate of drug excretion in urine against the plasma drug concentration. By analyzing the graph, the clearance can be calculated and obtained. Drugs rapidly excreted by the kidneys exhibit a...
High-Performance Liquid Chromatography: Elution Process