Variability of two methods of measuring absorbed dose in dental radiography
Abstract:
Thermoluminescence dosimetry (TLD) has been widely used in the monitoring of radiation exposure in healthcare workers operating radiation devices. This study compared the reliability of TLD to an electronic dosimeter. The absorbed dose registered by the TLD chips was found to be higher than that by the electronic dosimeter. Variability was found to be significantly higher and the variance component attributed to the variability of the method of measurement was more than twice as much in TLD when compared to the electronic dosimeter. It is concluded that TLD is a more variable method of measuring absorbed dose than the electronic dosimeter.
Related Concept Videos
Biological Effects of Radiation
X-ray Imaging
One-Compartment Open Model: Wagner-Nelson and Loo Riegelman Method for ka Estimation
On...
Measurement of Bioavailability: Pharmacodynamic Methods
Dosage Regimen: Individualization
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...