Routine quality control of recycled target [18O]water by capillary electrophoresis and gas chromatography
Bill X Huang1, Michael A Channing, Paul S Plascjak
1PET Department, Warren G. Magnuson Clinical Center, National Institutes of Health, Bethesda, MD 20892, USA. bhuang@mail.cc.nih.gov
Abstract:
Recycling of [(18)O]water for [(18)F]fluoride production can be accomplished with reliable results. We have developed sensitive, robust, and rapid analyses of impurities in [(18)O]water. Anions were quantitated by capillary electrophoresis and organic residuals were quantitated by gas chromatography using methods with excellent reproducibility and linearity. Kryptofix 222 (K-222) was quantitated by a sensitive LC-MS-MS technique. Isotopic composition was determined by GC-MS with satisfactory accuracy and precision. These methods were employed to evaluate recovered [(18)O]water purified by a novel electrolysis method. 2-[(18)F]FDG yields using purified [(18)O]water with very low levels of impurities are indistinguishable from newly purchased [(18)O]water. High (> 300 ppm) carbonate concentration reduces the fluoride trapping efficiency of QMA. The analyses of anions, organics, and isotopic enrichment were applied routinely for quality control of [(18)O]water to predict a satisfactory outcome of 2-[(18)F]FDG production.
Related Concept Videos
Electrophoresis: Overview
There...
Capillary Electrophoresis: Instrumentation
Capillary Electrophoresis: Applications
Capillary zone electrophoresis (CZE) separates ionic components based on their electrophoretic mobility. It has been used to separate proteins, amino acids,...