The effect of source to image distance on radiation risk to the patient
1UNITEC Institute of Technology, Auckland, New Zealand. jpoletti@unitec.ac.nz
Abstract:
The Monte Carlo package PCXMC was used to determine the effect of source to image distance (SID) on effective dose to the patient in diagnostic radiology. A range of common examinations was used, with a range of distances from 75 cm to 200 cm. The results showed that the radiation risk (effective dose) is virtually independent of the SID. In cases where a radiation sensitive organ is close to the edge of the x-ray field, the effective dose increases slightly with distance. In other cases there was either no dependence, or a small reduction in risk with increasing SID. There is therefore no justification in terms of radiation risk for a change in the standard SID of 1 m that is currently used for most projections.
Related Concept Videos
Biological Effects of Radiation
X-ray Imaging
Radiological Investigation I: X-ray and CT
Radiological Investigation III: Pulmonary Angiogram and PET Scan
Pulmonary Angiogram
A Pulmonary Angiogram is an invasive procedure involving injecting a contrast medium through a catheter threaded into the pulmonary artery or the right side of the heart to visualize the pulmonary vasculature. Computed Tomography (CT) scans have mainly replaced this...
Imaging Studies for Cardiovascular System III: X-Ray
Definition and Purpose
An X-ray, or radiograph, is a non-invasive method that uses ionizing radiation to take images of internal structures. It is mainly used in cardiac imaging to examine the heart, lungs, and major blood vessels, aiming to identify abnormalities in the heart's size, shape, and position, such as heart failure, congenital defects, and vascular...