Analysis of low levels of oxygen, carbon monoxide, and carbon dioxide in polyolefin feed streams using a pulsed
David B Wurm1, Kefu Sun, William L Winniford
1The Dow Chemical Company, 2301 Brazosport Blvd., Freeport, TX 77541, USA.
Abstract:
A gas chromatography (GC) method is developed for rapid analysis of polyolefin feed streams for the catalyst poisons CO, CO(2), and O(2). The method uses an HP MoleSieve column in parallel with a CP-PoraPLOT Q column and a pulsed discharge detector (PDD). Detection limits for each of the potential poisons are between 50 and 250 ppb. For a 10-ppm standard, the precision of the method was +/- 4.2% for oxygen, +/- 7.8% for carbon dioxide, and +/- 2.0% for carbon monoxide. In addition to the polyolefin feed stream, nitrogen and hydrogen feed streams are also analyzed. In each case, sampling is observed to be a critical issue, with air contamination of the sample cylinder often the limiting step in determining the true level of oxygen. It is also noted that large amounts of argon are present in the standards when nitrogen is used as a balance gas. Because the trace oxygen peak partly coelutes with the larger argon peak, it is suggested that helium be used as the balance gas for all standards. This general experimental arrangement should be effective when applied to feed streams for other polymers as well.
Related Concept Videos
Gas Chromatography: Types of Detectors-II
Gas Chromatography: Types of Detectors-I
TCD is the earliest and most widely used detector that operates by measuring the changes in the thermal conductivity of the carrier gas. When a sample compound enters the detector,...
Gas Chromatography: Overview of Detectors
A non-destructive detector allows a sample to be analyzed without altering or consuming it, meaning the sample can be collected after detection for further analysis. Examples include thermal conductivity detectors and...
High-Performance Liquid Chromatography: Types of Detectors
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...