Stopped flow apparatus for time-resolved Fourier transform infrared difference spectroscopy of biological
1Micro-biolytics GmbH, Georges Koehler Allee 102, D-79110 Freiburg, Germany.
Abstract:
Stopped flow spectroscopy is an established technique for acquiring kinetic data on dynamic processes in chemical and biochemical reactions, and Fourier transform infrared (FT-IR) techniques can provide particularly rich structural information on biological macromolecules. However, it is a considerable challenge to design an FT-IR stopped flow system with an optical path length low enough for work with aqueous (1H2O) solutions. The system presented here is designed for minimal sample volumes (approximately 5 microL) and allows simultaneous FT-IR rapid-scan and VIS measurements. The system employs a micro-structured diffusional mixer to achieve effective mixing on the millisecond time scale under moderate flow and pressure conditions, allowing measurements in a cell path length of less than 10 microns. This makes it possible to record spectra in 1H2O solutions over a wide spectral range. The system layout is also designed for a combination of kinetic and static measurements, in particular to obtain detailed information on the faster spectral changes occurring during the system dead time. A detailed characterization of the FT-IR stopped flow system is presented, including a demonstration of the alkaline conformational transition of cytochrome c as an example.