Related Experiment Videos
Flow artifacts in steady-state free precession cine imaging.
Pippa Storey1, Wei Li, Qun Chen
1Radiology Department, Evanston Hospital, Evanston Northwestern Healthcare and Northwestern University Feinberg School of Medicine, Evanston, Illinois 60201, USA. p-storey@northwestern.edu
Magnetic Resonance in Medicine
|January 6, 2004
Summary
Dark flow artifacts in cardiac cine imaging can be caused by incorrect scanner tuning. This study explains these artifacts and suggests tuning to the water peak, not the lipid peak, to prevent them.
Area of Science:
- Medical Imaging
- Magnetic Resonance Imaging Physics
Background:
- Steady-state free precession (SSFP) cardiac cine imaging is susceptible to dark flow artifacts.
- These artifacts often degrade image quality but can typically be resolved through scanner adjustments.
Purpose of the Study:
- To provide a theoretical explanation for dark flow artifacts in SSFP cardiac cine imaging.
- To validate the theory using phantom experiments and in vivo observations.
- To identify the likely cause of these artifacts in clinical settings.
Main Methods:
- Developing a theoretical model for dark flow artifacts based on off-resonance spin behavior.
- Conducting phantom experiments to replicate and validate the theoretical predictions.
- In vivo experiments involving controlled detuning of the scanner's center frequency.
Main Results:
- The theoretical model successfully explains dark flow artifacts as spins traversing off-resonant magnetic field points.
- Phantom experiments validated the theoretical predictions.
- In vivo artifacts were reproducible by detuning the center frequency by approximately half the inverse repetition time (TR).
- This detuning magnitude is comparable to the water-lipid frequency difference, suggesting mistuning to the lipid peak.
Conclusions:
- Dark flow artifacts in SSFP cardiac imaging arise from spins encountering off-resonant magnetic fields.
- Incorrectly tuning the scanner's center frequency to the lipid peak instead of the water peak is a probable cause of these artifacts in vivo.
- Proper scanner tuning is crucial for artifact-free cardiac cine imaging.