Quantitative vibration threshold testing in carpal tunnel syndrome: analysis strategies for optimizing reliability
Mark C Hubbard1, Joy C MacDermid, John F Kramer
1Orthopaedic & Rehab Department, Penn State Hershey Medical Center, Hershey, Pennsylvania, USA.
Summary
A new computer-controlled vibrometer offers a reliable method for measuring vibration sensation thresholds, crucial for early carpal tunnel syndrome (CTS) detection. This tool provides a quantitative and comfortable assessment of median nerve function in CTS patients.
Area of Science:
- Neurology
- Biomedical Engineering
Background:
- Vibration sensation thresholds are affected early in carpal tunnel syndrome (CTS).
- Traditional methods like tuning forks and electronic vibrometers have limitations in quantifying these thresholds.
- A novel computer-controlled vibrometer (PCV50) was developed to address these limitations.
Purpose of the Study:
- To establish a reliable testing procedure for a new computer-controlled vibrometer (PCV50).
- To assess the vibrometer's utility in quantifying vibration sensation thresholds in patients with carpal tunnel syndrome.
Main Methods:
- Fifty-two patients with confirmed carpal tunnel syndrome underwent testing with the PCV50 at a fixed frequency of 50 Hz.
- Three trials were conducted on the third digit of each hand, followed by a retest session 40-60 minutes later.
- Intraclass correlation coefficients (ICCs) were used to evaluate different data analysis strategies for reliability.
Main Results:
- The most reliable testing strategy involved averaging the second and third trials and comparing this to the retest score.
- This strategy yielded high intraclass correlation coefficients (ICCs) of 0.86 for the right hand and 0.89 for the left hand.
- The computer-controlled vibrometer demonstrated ease of administration, quantification, and patient comfort.
Conclusions:
- A reliable testing procedure was identified for the computer-controlled vibrometer.
- This validated procedure facilitates quantitative assessment of median nerve function.
- The PCV50 shows promise for early detection and accurate quantification of carpal tunnel syndrome impairments.