High-performance liquid chromatographic method for quantitating plasma levels of amiloride and its analogues
M A Alliegro1, K D Dyer, E J Cragoe
1Retina Center, St. Joseph Hospital, Baltimore, MD 21284.
Abstract:
An assay for amiloride was devised for efficient use with the wide variety of analogues available. Amiloride was extracted from 1-ml plasma samples by elution from a C8 preparative column with 6% acetonitrile-45% methanol-5.4% acetic acid, adjusted to pH 4.0 with trimethylamine. Samples were lyophilized, resuspended in 50% methanol, filtered through 0.22-microns Spin-X cartridges, applied to a reversed-phase C18 column, and eluted in a 0-50% acetonitrile gradient in 0.4% acetic acid, pH 4.5 (1.2 ml/min). Detection by ultraviolet absorbance at 360 nm was linear from 1 to 1000 ng. Versatility of the method was demonstrated with the analogues benzamil, 6-hydro-, 6-iodo-, 5-hexamethylene-, and 5-chlorobenzyl-2',4'-dimethylbenzyl-amiloride.