A New Hybrid Quantitative Evaluation Model for Axillary Junctional Hemorrhage in Swine
Published on: December 6, 2024
Venous blood redistribution alters the accuracy of Omron HEM-705CP
Marek Czarkowski1, Małgorzata Mikulska, Mariusz ZEbrowski
1Department of Internal Medicine and Endocrinology, Medical University of Warsaw, Warsaw, Poland. mczark@amwaw.edu.pl
Background:
Validation procedures of automated blood pressure measuring devices are carried out at rest only.
Design:
We aimed to determine whether venous blood redistribution affects the accuracy of the OMRON HEM-705CP (Osc). Lower body negative pressure stimulus (-40 mmHg) (LBNP) was used to simulate changes of body position.
Methods:
Fifty-four young healthy volunteers had their blood pressure (BP) measured simultaneously by Osc and mercury sphygmomanometer (HgS) at rest (at 3 and 5 min) and during LBNP (at 7 and 9 min).
Results:
Differences (delta) between HgS and Osc for systolic and diastolic BP (SBP and DBP) were calculated. The deltaSBP obtained in minute 9 (deltaSBP-9) was bigger than the one taken in minute 5(deltaSBP-5) (-2.89+/- 4.3 versus -1.35+/-5.2 mmHg, P<0.02). The rest of the deltaSBP values and all the deltaDBP values were constant at rest and during venous blood redistribution caused by LBNP. We calculated the fall in SBP separately for HgS and Osc by subtracting SBP-9 from SBP-5 for each patient. The fall measured by Osc was bigger than by HgS (-5.87+/-4.0 versus -4.3+/-4.5 mmHg; P<0.02). When graded according to British Hypertension Society protocol, Osc remained in the class of recommendation (A or B) for every minute of the procedure.
Conclusions:
Even if an oscillometric device passes various validation tests, they do not give sufficient information about its accuracy during venous blood redistribution. This knowledge is of particular importance when BP regulation abnormalities occurring during changes in body position are diagnosed with the oscillometric method.
Related Concept Videos
Pre-Procedural Guidelines for Assessing Blood Pressure
Errors occurring during blood pressure monitoring
Several factors...
Assessing Blood pressure using a doppler ultrasound
Pre-Procedural Guidelines for Doppler Ultrasound Blood Pressure Assessment:
Preparation of Equipment:
Special considerations while measuring blood pressure
Monitoring Both Arms:
Monitoring BP in both arms during the initial assessment is advisable, as the systolic value may differ by five to ten mm Hg between arms. For subsequent BP assessments, use the arm with the higher reading.
Assessing Blood pressure in the Leg
Preparation:
Venous Thrombosis IV: Nursing Management