Determination of zinc by flow injection with fluorimetric detection in a micellar medium
N G Gutiérrez1, F S Rojas, J M Pavón
1Department of Analytical Chemistry, Faculty of Sciences, University of Málaga, E-29071, Málaga, Spain.
Abstract:
A room-temperature flow injection spectrofluorimetric method is presented for the determination of Zn(II), based on the use of salicylaldehyde thiocarbohydrazone in the presence of Triton X-100 and sodium acetate-acetic acid buffer. Various physical and chemical variables affecting the reaction in the flow system were evaluated. The proposed method is very selective. The calibration graph is linear over the range 10-1000 ng/ml, with a detection limit of 5 ng/ml and a relative standard deviation at the 50 ng/ml level of 1.8 %. The method was successfully applied to the determination of Zn(II) in drinking waters and biological samples.