Related Experiment Video
Updated: Aug 24, 2026

Three-Dimensionally Printed Microfluidic Cross-flow System for Ultrafiltration/Nanofiltration Membrane Performance Testing
Published on: February 13, 2016
In situ fabricated porous filters for microsystems
Jaisree Moorthy1, David J Beebe
1Department of Biomedical Engineering, University of Wisconsin at Madison, WI 53706, USA.
Abstract:
Porous filters were fabricated inside a microchannel using emulsion photo-polymerization. The porosity of the filter is dependent on the composition of the pre-polymer mixture. The morphology and flow resistance of four filters with varying water and cross-linker concentrations was studied. The porous filter was used to separate samples, such as whole blood to cell/serum components, based on size. The efficiency of cell/serum separation by the porous filter was tested by performing a glucose-6-phosphate dehydrogenase assay and was found to be comparable to separation by centrifuge. The porous filter thus mimics the functionality of the centrifuge, with the added benefits of no power requirements and ability to handle small sample volumes.