Determination of a cyclooxygenase II inhibitor in human plasma by capillary gas chromatography with mass spectrometric detection
Related Experiment Video
Updated: Aug 11, 2026

Mass Spectrometry and Luminogenic-based Approaches to Characterize Phase I Metabolic Competency of In Vitro Cell Cultures
Published on: March 28, 2017
J D-Y Dru1, C M Chavez-Eng, M L Constanzer
1Department of Drug Metabolism, Merck Research Laboratories, Merck & Co. Inc., P.O. Box 4, West Point, PA 19486, USA. james_dru@merck.com
Abstract:
Sensitive methods based on capillary gas chromatography (GC) with mass spectrometric (MS) detection in a selected-ion monitoring mode (SIM) for the determination of a cyclooxygenase II (COX-II) inhibitor (3-isopropoxy-4-(4-methanesulfonylphenyl)-5,5'-dimethyl-5H-furan-2-one, I) in human plasma, in two concentration ranges of 0.1-20 and 5-1000 ng/ml, are described. Following liquid-liquid extraction, the residue, after evaporation of the organic phase to dryness, was reconstituted in acetonitrile (20 l) and part of the extract (1 l) was analyzed by GC/MS/SIM. The drug (I) and internal standard (II) were separated on a 25 mx0.2 mm capillary column with HP Ultra 1 (100% dimethylpolysiloxane, 0.33 m) phase and analyzed by MS/SIM monitoring ions at m/z 237 and 282 for I and II, respectively. The standard curve was linear within the lower concentration range of 0.1-20 ng/ml and the lower limit of quantification (LLOQ) in plasma was 0.1 ng/ml. Intraday coefficients of variation (CV, n=5) were 8.9, 4.2, 5.7, 3.1, 1.9, 1.9, and 4.4% at 0.1, 0.2, 0.5, 1.0, 5.0, 10, and 20 ng/ml, respectively. The standard curve was also linear within the higher concentration range of 5-1000 ng/ml and the LLOQ in plasma was 5 ng/ml. Intraday coefficients of variation (CV, n=5) were all below 9% at all concentrations within the standard curve range. The accuracy for I in human plasma was 91-112% and the recovery of I and II was greater than 70% at all concentrations within both standard curve ranges. The details of the assay methodology are presented.
More Related Videos
10:17High-throughput and Comprehensive Drug Surveillance Using Multisegment Injection-Capillary Electrophoresis-Mass Spectrometry
Published on: April 23, 2019
07:14Detection and Quantification of Calcitonin Gene-Related Peptide (CGRP) in Human Plasma Using a Modified Enzyme-Linked Immunosorbent Assay
Published on: June 16, 2023
Related Concept Videos
High-Resolution Mass Spectrometry (HRMS)
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Inductively Coupled Plasma–Mass Spectrometry (ICP–MS): Overview
Gas Chromatography: Types of Detectors-II