Background:
This study determines the analytical characteristics of the i-STAT cardiac troponin I assay (cTnI; i-STAT, Princeton, NJ), a 10-min POC assay, designed to be performed at the bedside.
Methods:
Three different hospitals participated in a patient specimen and analytical validation study (n=186) for the i-STAT cTnI assay carried out in real time. A total of 186 whole blood specimens (lithium heparin) were collected from patients presenting with symptoms suggestive of acute coronary syndromes (ACS) for correlation studies as well as from 162 healthy subjects for reference interval determination. Factors studied included antibody specificity, detection limit, imprecision, linearity, assay specificity, sample type stability, interferences, reference limit determination and comparison vs. the Dade Stratus CS cTnI assay.
Results:
Total imprecision (CV) of 10% and 20% were seen at 0.09 and 0.07 microg/l, respectively. The detection limit was 0.02 microg/l. The 99th percentile reference limit was 0.08 microg/l. The assay was not affected by common interferents. An equimolar response within 5% was found for reduced and phosphorylated forms of TIC and IC complexes. Regression analysis for the i-STAT cTnI between whole blood and plasma specimens and for whole blood between the i-STAT and Stratus CS cTnI assays demonstrated slopes of 1.06 and 0.89, respectively.
Conclusions:
The i-STAT cTnI assay is a sensitive and precise monitor of cTnI, poised for point-of-care/near bedside clinical utilization for triage, diagnostics and risk management of acute coronary syndrome patients.
More Related Videos
07:40Impact of High-intensity Interval Exercise and Moderate-Intensity Continuous Exercise on the Cardiac Troponin T Level at an Early Stage of Training
Published on: October 10, 2019
09:23Using Extraordinary Optical Transmission to Quantify Cardiac Biomarkers in Human Serum
Published on: December 13, 2017
Related Concept Videos
Blood Studies for Cardiovascular System I: Cardiac Biomarkers
The essential diagnostic tools for detecting myocardial necrosis and monitoring individuals suspected of having acute coronary syndrome (ACS) include:
Troponins
Troponins, particularly cardiac troponins I and T, are the most precise and sensitive markers of myocardial injury. They are detectable within 4-6 hours of myocardial injury and remain...
Acute Coronary Syndrome III: Diagnostic Studies
Acute Coronary Syndrome I: Introduction
Myocarditis II: Clinical Features and Diagnostic Tests