Related Experiment Video
Updated: Aug 23, 2026

Comprehensive Characterization of Extended Defects in Semiconductor Materials by a Scanning Electron Microscope
Published on: May 28, 2016
An ultrahigh-vacuum goniometer for in situ soft X-ray standing-wave analysis of semiconductor surfaces
1NTT Basic Research Laboratories, 3-1 Morinosato, Wakamiya, Atsugi-shi, Kanagawa 243-01, Japan.
Abstract:
An ultrahigh-vacuum goniometer was developed for in situ X-ray standing-wave (XSW) analysis of semiconductor surfaces prepared by molecular-beam epitaxy (MBE). Although two ultrahigh-vacuum motors for chi and phi rotating axes are inside the analysis chamber, low-energy photoelectrons can still be collected as the magnetic field is sufficiently suppressed by using metal shields. Furthermore, the sample can be annealed at temperatures higher than 870 K on the goniometer in the analysis chamber. This goniometer is used at beamline 1A (BL-1A) at the Photon Factory, where both monochromated soft X-rays and UV radiation are available. This analysis system was shown to be suitable not only for in situ soft-XSW and X-ray absorption near-edge structure (XANES) studies but also for synchrotron radiation photoelectron spectroscopy (SRPES) studies. The annealing effects on an S-adsorbed GaAs(001) surface could be studied by SRPES, XANES and XSW using this new goniometer.