Unexpected loss of atrial tracking caused by interaction between temporary and permanent right ventricular leads
Kuan-Cheng Chang1, Jan-Yow Chen, Jen-Jyh Lin
1Division of Cardiology, Department of Internal Medicine, China Medical University Hospital, Taichung, Taiwan. kcc45@ms36.hinet.net
Abstract:
This report describes a patient who underwent implantation of an atriobiventricular pacemaker following AV junction ablation and insertion of a temporary right ventricular (RV) pacemaker. During implantation, intermittent loss of sinus P wave tracking occurred when the three permanent leads were connected to the generator. Analysis of marker annotation disclosed intermittent abnormal ventricular sensing that reinitiated postventricular atrial blanking and caused failure of P wave tracking. This phenomenon disappeared after removing the temporary RV lead, but not by turning off the temporary pacemaker. We assume that mechanical contact between the temporary and the permanent RV leads is the underlying mechanism.
More Related Videos
16:40A New Single Chamber Implantable Defibrillator with Atrial Sensing: A Practical Demonstration of Sensing and Ease of Implantation
Published on: February 28, 2012
12:45Benefits of Cardiac Resynchronization Therapy in an Asynchronous Heart Failure Model Induced by Left Bundle Branch Ablation and Rapid Pacing
Published on: December 11, 2017
Related Concept Videos
Disturbances in Heart Rhythm
Arrhythmias are categorized by their speed, rhythm, and origin. A slow heart...
Dysrhythmias IV: Characteristics of Bradyarrhythmias
ECG Interpretation of Arrhythmias II: Atrial, Junctional and Ventricular Arrhythmias