Preparation and 3D Tracking of Catalytic Swimming Devices
Published on: July 1, 2016
Determination and correction of position detection nonlinearity in single particle tracking and three-dimensional
Christian Tischer1, Arnd Pralle, Ernst-Ludwig Florin
1Cell Biology and Biophysics Programme, European Molecular Biology Laboratory, Meyerhofstrasse 1, 69117 Heidelberg, Germany.
Abstract:
A general method is presented for determining and correcting nonlinear position detector responses in single particle tracking as used in three-dimensional scanning probe microscopy based on optical tweezers. The method uses locally calculated mean square displacements of a Brownian particle to detect spatial changes in the sensitivity of the detector. The method is applied to an optical tweezers setup, where the position fluctuations of a microsphere within the optical trap are measured by an interferometric detection scheme with nanometer precision and microsecond temporal resolution. Detector sensitivity profiles were measured at arbitrary positions in solution with a resolution of approximately 6 nm and 20 nm in the lateral and axial directions, respectively. Local detector sensitivities are used to reconstruct the real positions of the particle from the measured position signals.