Related Experiment Video
Updated: Aug 11, 2026

Dynamic Lung Tumor Tracking for Stereotactic Ablative Body Radiation Therapy
Published on: June 7, 2015
Intensity-modulated radiation therapy: overlapping co-axial modulated fields
P Metcalfe1, P Tangboonduangjit, P White
1Centre for Medical Radiation Physics, University of Wollongong, NSW, Australia.
Abstract:
The Varian multi-leaf collimator has a 14.5 cm leaf extension limit from each carriage. This means the target volumes in the head and neck region are sometimes too wide for standard width-modulated fields to provide adequate dose coverage. A solution is to set up asymmetric co-axial overlapping fields. This protects the MLC carriage while in return the MLC provides modulated dose blending in the field overlap region. Planar dose maps for coincident fields from the Pinnacle radiotherapy treatment planning system are compared with planar dose maps reconstructed from radiographic film and electronic portal images. The film and portal images show small leaf-jaw matchlines at each field overlap border. Linear profiles taken across each image show that the observed leaf-jaw matchlines from the accelerator images are not accounted for by the treatment planning system. Dose difference between film reconstructed electronic portal images and planning system are about 2.5 cGy in a modulated field at d(max). While the magnitude of the dose differences are small improved round end leaf modelling combined with a finer dose calculation grid may minimize the discrepancy between calculated and delivered dose.

