A new mechanical testing device for measuring anteroposterior knee laxity
Andreas J Schuster1, Mike J McNicholas, Stefan W Wachtl
1Department of Orthopaedic Surgery, Hôpital Cantonal, Fribourg, Switzerland. andreas.schuster@spitalbern.ch <andreas.schuster@spitalbern.ch>
The American Journal of Sports Medicine
|October 21, 2004
Summary
The Rolimeter is a valid, reproducible, and reliable instrument for measuring knee anteroposterior translation, comparable to the KT-1000 arthrometer. It offers accurate knee laxity measurements.
Area of Science:
- Orthopedic biomechanics
- Medical device evaluation
- Knee joint kinematics
Background:
- A novel instrument, the Rolimeter, was evaluated for measuring anteroposterior (AP) translation in the knee.
- The Rolimeter's performance was benchmarked against the established KT-1000 arthrometer.
Purpose of the Study:
- To ascertain if the Rolimeter provides a valid, reproducible, and reliable method for assessing knee AP translation.
- To compare the Rolimeter's efficacy against the KT-1000 arthrometer.
Main Methods:
- Two observers assessed 16 healthy subjects (32 knees) and 36 patients (72 knees) with ligament ruptures.
- Measurements of total AP translation were taken twice, 30 minutes apart, using both the Rolimeter and KT-1000 at 20 and 80 degrees of knee flexion during a manual maximal Lachman test.
Main Results:
- The Rolimeter consistently measured approximately 1 mm less AP displacement than the KT-1000 under maximal manual stress.
- Rolimeter measurements demonstrated superior consistency across observers compared to the KT-1000.
- Both instruments exhibited equivalent sensitivity and specificity for detecting knee instability.
Conclusions:
- The Rolimeter is confirmed as a reproducible and reliable tool for measuring knee AP translation, on par with the KT-1000 arthrometer.
- The Rolimeter is a valid instrument for quantifying knee AP translation.
- Optimal accuracy for the KT-1000 was observed at 20 degrees flexion, while the Rolimeter showed higher accuracy at 80 degrees flexion.