Phantom design method for high-field MRI human systems.

Qing X Yang1, Jinghua Wang, Christopher M Collins

  • 1Center for NMR Research, Department of Radiology, Pennsylvania State University College of Medicine, Hershey 17033, USA. qyang@psu.edu

Magnetic Resonance in Medicine
|October 28, 2004
PubMed
Summary
This summary is machine-generated.

Related Concept Videos

You might also read

Related Articles

Articles linked to this work by shared authors, journal, and citation graph.

Sort by
Same author

First-Order Spatial Encoding Simulations for Improved Accuracy in the Presence of Strong B<sub>0</sub> and Gradient Field Variations.

Magnetic resonance in medicine·2025
Same author

Multiple cation insertion into a polyaromatic hydrocarbon guided by data and computation.

Chemical science·2025
Same author

Discovering trends in big data: general discussion.

Faraday discussions·2024
Same author

Discovering structure-property correlations: general discussion.

Faraday discussions·2024
Same author

Discovering chemical structure: general discussion.

Faraday discussions·2024
Same author

Exploration of Chemical Space Through Automated Reasoning.

Angewandte Chemie (International ed. in English)·2024
Same journal

Suppression of Oscillation and Ghosting in RF-Spoiled Gradient-Echo-Based Dynamic Imaging.

Magnetic resonance in medicine·2026
Same journal

A Simple, Dynamic Geometric Phantom for MRI and CT Reconstruction Pipelines: Beyond Shepp-Logan.

Magnetic resonance in medicine·2026
Same journal

7T 3D-EPI PCASL With High SNR Efficiency and Robustness to Through-Plane B<sub>0</sub> Field Gradients.

Magnetic resonance in medicine·2026
Same journal

A Comparison of Tissue Property Values Estimated Using Conventional Cardiac MRF and MT-Cardiac MRF.

Magnetic resonance in medicine·2026
Same journal

Dependence of the Extra-Cellular Diffusion Coefficient on the Fractions of Neurites and Cell Bodies in Gray Matter.

Magnetic resonance in medicine·2026
Same journal

Triple-Pulse <sup>23</sup>Na MRI Sequence (TriNa) for Simultaneous Acquisition of Spin-Density-Weighted and Fluid-Attenuated Images.

Magnetic resonance in medicine·2026
See all related articles

This study presents a novel phantom design for high-field magnetic resonance imaging (MRI) that mimics radiofrequency (RF) field wave characteristics. This method accurately reproduces human sample image patterns across different field strengths.

Area of Science:

  • Magnetic Resonance Imaging (MRI)
  • Biophysics
  • Medical Imaging Physics

Background:

  • High-field MRI (7.0 T and 3.0 T) presents challenges in RF field homogeneity.
  • Understanding RF field wave characteristics is crucial for accurate phantom design.
  • Existing phantoms may not fully replicate complex wave behaviors observed in human samples.

Purpose of the Study:

  • To present a phantom design method for high-field MRI.
  • To validate the method using experimental data at 7.0 T and 3.0 T.
  • To demonstrate the reproduction of RF field wave characteristics in phantoms.

Main Methods:

  • The phantom design is based on the RF field wave characteristics of the sample.
  • Key factors considered are sample size relative to wavelength and current ratios.

Related Experiment Videos

  • Experimental validations were performed at 7.0 T and 3.0 T.
  • Main Results:

    • RF field distribution is primarily determined by sample dimensions and current properties.
    • MR image intensity patterns mimicking wave behavior were successfully reproduced.
    • The phantom design allows for scaling to match different field strengths and sample media.

    Conclusions:

    • A scalable phantom design method for high-field MRI is established.
    • The method effectively replicates RF field wave characteristics observed in human samples.
    • This approach enhances the utility of phantoms for MRI research and development.