Matrix media selection for the determination of residual solvents in pharmaceuticals by static headspace gas
Koji Urakami1, Atsuya Higashi, Kazuichi Umemoto
1Analytical Development Laboratories, Takeda Pharmaceutical Company Limited, 2-17-85 Jusohonmachi, Yodogawa-ku, Osaka 532-8686, Japan. urakami_koji@takeda.co.jp
Abstract:
The influence of the matrix medium used for the determination of residual solvents in pharmaceuticals by static headspace gas chromatography was investigated. The purpose of this paper is to propose a guide for the choice of a matrix medium suitable for the determination of residual solvents of interest. Dimethylsulfoxide (DMSO), N,N-dimethyformamide (DMF), N,N-dimethylacetamide (DMA), benzyl alcohol (BA), 1,3-dimethyl-2-imidazolidinone (DMI) and water were studied as matrix media, and seventeen solvents used for the synthesis and purification of drug substances were used as target analytes. The peak shape of each analyte was not affected by the matrix medium, whereas the peak intensities for all solvents were strongly affected by the matrix medium; for example, the peak intensity of methanol in a BA matrix was more than four times that in a DMSO matrix. With a few exceptions, the peak intensities are approximately doubled for every 20 degrees C rise in equilibrium temperature between 80 and 140 degrees C, and there is no difference in this behavior among the matrix media. In addition, the formation of artifacts from the matrix media, upon heating in a headspace sampling apparatus, was investigated. Artifacts were also formed following ultrasonication of sample solutions used to increase dissolution of the sample into the matrix medium selected. These artifacts included benzene and toluene which were restricted as Class 1 and 2 toxic solvents in the ICH guideline.
More Related Videos
08:56Detection of Regulated Ergot Alkaloids in Food Matrices by Liquid Chromatography-Trapped Ion Mobility Spectrometry-Time-of-Flight Mass Spectrometry
Published on: November 22, 2024
04:19An Improved Technique for Trimethylamine Detection in Animal-Derived Medicine by Headspace Gas Chromatography-Tandem Quadrupole Mass Spectrometry
Published on: March 10, 2023
Related Concept Videos
Gas Chromatography–Mass Spectrometry (GC–MS)
A gas chromatograph consists of a long, narrow capillary column with a polysiloxane coating on the inner wall. The coating...
Gas Chromatography: Introduction
In GC, a sample is vaporized and mixed with an inert carrier gas (the mobile phase), which transports it through a column.
Gas Chromatography: Types of Columns and Stationary Phases
For an analyte to remain on the column for a sufficient amount of time, it must exhibit some level of compatibility (or...
Supercritical Fluid Chromatography
SFC utilizes a supercritical fluid mobile phase,...
Chromatographic Methods: Classification
Chromatographic techniques are typically named by...
Chromatographic Methods: Terminology